4.6 Article

Role of Surgical Position on Interface Stress and Initial Bone Remodeling Stimulus around Hip Resurfacing Arthroplasty

期刊

JOURNAL OF ARTHROPLASTY
卷 24, 期 7, 页码 1137-1142

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2008.08.005

关键词

hip resurfacing; finite element model; stress shielding; surgical technique; bone remodeling

资金

  1. Stryker Orthopaedics

向作者/读者索取更多资源

Valgus alignment of femoral resurfacing components has been advocated to reduce proximal femur loading and thus minimize the risk for femoral neck fractures. However, Such reduction in loading may exacerbate undesirable stress shielding. This Study examined the effect of extreme implant orientations (+/- 50 degrees) and stem canal overreaming on initial bone remodeling stimulus using finite element models. The changes in implant-cement interface stresses due to implant alignment were also evaluated. The valgus model showed increased initial bone resorption stimulus, which extended distally and peripherally around the femoral neck. The peak implant-cement interface shear stress for the varus model was 10.9 MPa, exceeding the interface shear strength. Overreaming of the stem canal eliminated distal tip loading, but proximal stress shielding was Still unavoidable. These data show bone loading and interface fixation trends emanating from valgus and varus implant positions that will be of interest to practicing physicians.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据