Custom-fit total knee arthroplasty (OtisKnee) results in malalignment

期刊

JOURNAL OF ARTHROPLASTY
卷 23, 期 1, 页码 26-29

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2007.10.001

关键词

knee; total knee arthroplasty; custom guides; computer navigation

向作者/读者索取更多资源

We report on 4 patients who underwent total knee arthroplasty with OtisKnee system (OtisMed, Hayward, Calif). An image-free computer navigation system was used to evaluate the deformities and the recommended ruts. The recommended custom cuts were as follows: valgus/varus cuts on the femur (5.5 degrees valgus to 0.5 degrees varus) in reference to the mechanical axis, flexion cuts on the femur (4 degrees-9 degrees of flexion); femoral rotation was within 1 degrees of the epicondylar axis; valgus/varus cut on the tibia (3 degrees of valgus to 7.5 degrees of varus); tibial slope (5.5 degrees of anterior slope to 0.5 degrees of posterior slope). The custom OtisKnee system guides recommended alignment of the components that was more than 3 degrees off of mechanical axis. The potential for malalignment with this system places implants at high risk of early failure.