期刊
JOURNAL OF ANALYTICAL CHEMISTRY
卷 64, 期 12, 页码 1242-1246出版社
MAIK NAUKA/INTERPERIODICA/SPRINGER
DOI: 10.1134/S1061934809120077
关键词
-
资金
- Chemistry Department of Ondokuz Mayis University
A simple method for determination of common inorganic anions in mushroom samples has been developed by using suppressed ion chromatography with a pH detection unit. The detection unit which was constructed in such a way that practically no additional dispersion occurred consisted of a flow-through quinhydrone pH sensor and a small reference electrode. Chromatographic separation was performed in the order F(-), Cl(-), NO (2) (-) , Br(-), PO (4) (3-) , ClO (3) (-) , NO (3) (-) , and SO (4) (2-) , at room temperature by using Ion Pac AS 9-HC anion exchange column. Anion extracts from dried mushroom samples at room temperature were homogenized and filtered before injection. Under optimized analytical conditions, the detection limits of the method were between 2 x 10(-6) and 3 x 10(-4) M, depending on the anion studied. The results showed that the concentrations of fluoride and bromide in all mushroom samples were below their limit of detection. Nitrite was found to be the lowest abundant ion, while the most abundant ion was sulfate in all the mushroom samples studied.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐