Relative spectral response calibration of a spectrometer system for laser induced breakdown spectroscopy using the argon branching ratio method

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 29, 期 4, 页码 657-664

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c3ja50371b

关键词

-

资金

  1. NSF [CHE-0822469]

向作者/读者索取更多资源

This paper reports the use of the Ar branching ratio as an alternative means of calibrating the relative spectral response of a spectrometer/detector apparatus, with emphasis on its use in Laser Induced Breakdown Spectroscopy (LIBS) diagnostic experiments. The method is compared to the more conventional calibration routine using standard lamps. The argon method was implemented with an Ar-filled hollow cathode lamp and with a LIBS plasma in argon at different pressures. For the conventional calibration, a quartz tungsten halogen (QTH) lamp available in our laboratory has been used as a standard of spectral irradiance. The results obtained with the three methods are compared and discussed. It is shown that the spectral response data agree well with each other above 350 nm. At lower wavelengths, the data differ significantly. It is believed that this is due to the degradation of the calibration values of the lamp, which was not re-calibrated in a long time. It is concluded that the Ar-branching ratio method is a useful calibration technique for use with LIBS plasma diagnostic experiments.