期刊
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 29, 期 7, 页码 1252-1257出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/c4ja00097h
关键词
-
The characterization, sizing, and quantification of metal-based nanoparticles (NP) in a variety of matrices using single particle-inductively coupled plasma-mass spectrometry (SP-ICP-MS) is becoming increasingly popular due to the sensitive nature of the technique. Nanoparticle events in the plasma are less than 0.5 ms in duration; however current quadrupole-based ICP-MS instruments are limited to instrument dwell times in the millisecond range and have data acquisition overhead that adversely affects data quality. Novel instrument settings and data processing techniques can be used to explore the benefits of continuous data acquisition rates as fast as 105 Hz (or 10 ms dwell times). This paper provides data on the different effects data acquisition rate has on the quality of data that can be obtained by SP-ICP-MS. The effect of varying the dwell time and its influence on particle integration, particle counting, particle sizing, and background signal is discussed. This paper provides data on identifying the significant instrument settings and their implications on nanoparticle characterization.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据