4.7 Article

The effect of laser pulse duration on ICP-MS signal intensity, elemental fractionation, and detection limits in fs-LA-ICP-MS

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 28, 期 11, 页码 1781-1787

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c3ja50200g

关键词

-

资金

  1. U.S. Department of Energy, office of National Nuclear Security Administration [DE-NA0000463]

向作者/读者索取更多资源

We investigated the role that pulse width has on the analytical capabilities of laser ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS), specifically in the range of 40 fs to 300 ps. Three main results of LA-ICP-MS were examined: signal intensity or sensitivity, elemental fractionation or that the ablated aerosol is representative of the bulk, and detection limits of various elements in the samples. The samples used for this experiment were NIST glass standards, dielectrics that exhibit different ablation properties compared to metal targets. Our results demonstrate that in the range of 40 fs to 1 ps, negligible differences occur in signal intensity, elemental ratios, and detection limits. The U/Pb and U/Th ratios, which were examined to ensure limited fractionation, give comparable results at all pulse widths investigated.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据