期刊
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 28, 期 8, 页码 1320-1326出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/c3ja50030f
关键词
-
资金
- National Science Council of the Republic of China [NSC98-2113-M-037-009-MY3, NSC 99-2627-M-037-001]
In this study, an on-line polyoxometalate cluster (POM)/microfluidic (MF) separation system coupled with inductively coupled plasma mass spectrometry (ICP-MS) detection was developed for the determination of trivalent [Cr(III)] and hexavalent [Cr(VI)] chromium species in aqueous samples. In this system, a POM-immobilized MF device was used to separate the chromium species prior to their determination through ICP-MS. Several parameters were investigated to optimize the performance of the system: the adsorption pH, the adsorption and elution flow rates, the elution solution, the adsorption capacity, and the reaction chamber volume. Under the optimized conditions (carrier solution: deionized H2O; adsorption/elution flow rates: 0.2/1.2 ml L-1; reaction chamber volume: 30 mu L), the linearity (> 0.9987), precision (< 3.69%; n = 7), and accuracy (95.93-108.60%) of this analytical procedure were quite high. To confirm its practical utility, the novel POM-MF separation system was applied to the successful determination of chromium species in real-world samples (including spiked tap water).
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据