4.7 Article

Application of ion molecule reaction to eliminate WO interference on mercury determination in soil and sediment samples by ICP-MS

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 26, 期 6, 页码 1198-1203

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c1ja00005e

关键词

-

资金

  1. National Nature Science Foundation of China [40973021, 40821061]

向作者/读者索取更多资源

Mercury is subject to severe tungsten oxide (WO) interference in the direct ICP-MS analysis for soil or sediment samples. In this study, a method based on ion molecule reaction (IMR) was used to eliminate the WO spectral interference in a dynamic reaction cell (DRC). Interfered ion ((WO+)-W-202) was drastically oxidized to higher oxide ions ((WO2+)-W-218 or (WO3+)-W-234) by O-2 as the reaction gas in the DRC, while the analyte ion (Hg-202(+)) cannot react with O-2. Under the optimized O-2 flow rate (1.9 mL min(-1)) and DRC rejection parameter q (Rpq, 0.80), the background signal was reduced by up to 1000-fold at m/z 202 and the limit of quantitation (LOQ, 10 sigma) for Hg-202 was 1 ng g(-1). The proposed method was applied to direct determination of Hg in a series of soil and sediment standard reference materials (SRMs) and the satisfactory results indicate that it has great potential for the determination of trace or ultra-trace level mercury in various environmental or geological samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据