Validation of isotopic analysis of depleted, natural and enriched uranium using high resolution ICP-OES

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 26, 期 2, 页码 293-299

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c0ja00083c

关键词

-

向作者/读者索取更多资源

Analytical procedures for the reliable determination of uranium isotopes employing a commercially available high resolution (HR)-ICP-OES installed in a glove box were developed and validated. To this end, nine commonly used U emission lines were tested for their potential to provide isotopic splitting of the U signal, thus providing isotopic information. Among all emission lines tested, the wavelength region around 424.4 nm offered the largest spread between individual U signals, i.e. (233)U, (235)U, (236)U, and (238)U. This study focused specifically on the reliable determination of the abundances of (235)U at 424.412 nm and (238)U at 424.437 nm in samples containing depleted, natural and slightly enriched U, but also included specimens with rather high (up to similar to 90%) (235)U enrichments. Appropriate selection of the spectral background and accurate positioning of the peaks proved essential for obtaining reproducible U isotope ratios. The accuracy of the developed methodology was confirmed by the analysis of eight certified isotopic reference materials as well as eight reference samples that have been characterised in an accredited laboratory in-house using TIMS. In addition, this study highlights the potential of the developed HR-ICP-OES procedure for identifying nuclear sources of various U specimens via the analysis of (233)U (424.398 nm) and (236)U (424.423 nm). In contrast to the commonly used techniques, the methodology applied in this study does neither require the separation of U from the matrix prior to analysis nor the regular analysis of a reference sample with a known isotopic composition to correct for bias effects as it is the case for more sophisticated and precise mass spectrometric measurements. Nevertheless, HR-ICP-OES provides fit for purpose U isotopic information with reasonable accuracy (typically <1.5%) and precision (similar to 1%) within a few minutes. As such, HR-ICP-OES can be employed as a reliable, fast screening tool to identify the U isotopic composition and as such, complements the more laborious TIMS analysis routinely employed for this purpose.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据