Selective separation of selenium (IV) by thiol cellulose powder and subsequent selenium isotope ratio determination using multicollector inductively coupled plasma mass spectrometry

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 23, 期 5, 页码 733-743

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b801673a

关键词

-

向作者/读者索取更多资源

An efficient matrix separation procedure for the removal of hydride-forming elements such as As, Ge and Sb and transition metal ion interferences is described for the determination of selenium isotope ratio by hydride generation coupled to multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS). The preconcentration of Se ions by thiol cellulose powder (TCP) was studied by both batch and column methods. The selective separation of Se (IV) from the interferences was achieved with a column method using 0.10 g of TCP and eluting the interfering species with dilute hydrochloric acid prior to analysis. A recovery of >= 98 % was obtained using the column method. The procedure was also used to evaluate the degree of pre-concentration of Se on the TCP. The degree of pre-concentration was linearly proportional to the amount of sample passed through the column. Se retained on the TCP was quantitatively recovered by digesting twice with small volumes ( 50 mu L) of concentrated nitric acid. The complete elimination of interfering elements and the marked gain in sensitivity allowed high precision Se isotope ratios measurements in lake-sediments and certified reference materials using hydride generation (HG) coupled to MC-ICP-MS. The estimated external precision achieved for delta Se-78 (parts per thousand) by HG-MC-ICP-MS after sample pre-concentration and matrix removal with TCP columns varied between 0.15 parts per thousand and 0.21 parts per thousand using a double spike correction approach.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据