4.7 Article

Real-time multielement monitoring of airborne particulate matter using ICP-MS instrument equipped with gas converter apparatus

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 23, 期 8, 页码 1125-1129

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b802302f

关键词

-

向作者/读者索取更多资源

The real-time multielement monitoring of airborne particulate matter (APM) was carried out using an inductively coupled plasma mass spectrometry (ICP- MS) instrument equipped with a newly developed gas converter apparatus. By using the gas converter apparatus, gas molecules in the air sample and Ar molecules were almost completely exchanged and the gas- converted air sample could be directly introduced into ICP- MS instruments. Fe in clean room and outdoor air samples was directly measured using the ICP-MS instrument equipped with the converter apparatus. The signal intensities of Be, Ag, Cd, Sn, Sb, Tl, Pb, Bi, Th, and U in the outdoor air sample were successfully measured at every 8 min for 77 h.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据