期刊
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY
卷 57, 期 10, 页码 3974-3980出版社
AMER CHEMICAL SOC
DOI: 10.1021/jf900587m
关键词
Melamine detection; infant formula powder; infrared spectroscopy
资金
- U.S. Department of Agriculture [1935-42000-035]
Near- and mid-infrared spectroscopy methods (NIR, FTIR-ATR, FTIR-DRIFT) were evaluated for the detection and quantification of melamine in infant formula powder. Partial least-squares (PLS) models were established for correlating spectral data to melamine concentration: R-2 > 0.99, RMSECV <= 0.9, and RPD >= 12. Factorization analysis of spectra was able to differentiate unadulterated infant formula powder from samples containing 1 ppm melamine with no misclassifications, a confidence level of 99.99%, and selectivity > 2. These nondestructive methods require little or no sample preparation. The NIR method has an assay time of 1 min, and a 2 min total time to detection. The FTIR methods require up to 5 min for melamine detection. Therefore, NIR and FTIR methods enable rapid detection of 1 ppm melamine in infant formula powder.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。