In Vitro Tests for Aerosol Deposition. III: Effect of Inhaler Insertion Angle on Aerosol Deposition

出版社

MARY ANN LIEBERT, INC
DOI: 10.1089/jamp.2012.0989

关键词

Mouth-throat airway model; metered dose inhaler; dry powder inhaler; insertion angle; in vitro deposition; CFD modeling

资金

  1. Medical College of Virginia Foundation

向作者/读者索取更多资源

Background: Inhaler orientation with respect to a patient's mouth may be an important variable determining the efficiency of aerosol lung delivery. The effect of insertion angle on regional deposition was evaluated for a series of inhalers using concurrent in vitro and computational fluid dynamics (CFD) analysis. Methods: Geometrically realistic physical mouth-throat (MT) and upper tracheobronchial (TB) models were constructed to connect different inhalers at a series of insertion angles relative to the horizontal plane of the model. These models were used to assess albuterol sulfate deposition from the Novolizer (R) dry powder inhaler (DPI), Proventil (R) HFA pressurized metered dose inhaler (MDI), and Respimat (R) Soft Mist (R) Inhaler (SMI) following the actuation of a single dose. Drug deposition from Novolizer DPI was studied for Salbulin (R) and an experimental drug only formulation. Albuterol sulfate was recovered and quantified from the device and the MT and TB regions. Results: Significant differences in MT and total lung dose (TLD) of albuterol sulfate deposition were not observed for Salbulin Novolizer DPI and Respimat SMI inserted at different angles. In contrast, drug-only Novolizer DPI and Proventil HFA MDI showed a significant difference in MT and TLD deposition using different insertion angles. For drug-only Novolizer DPI and Proventil HFA MDI, the lowest and the highest MT depositions were observed at +10 degrees and -20 degrees, respectively; for Respimat SMI and Salbulin Novolizer DPI, these angles were -10 degrees and +10 degrees, and +20 degrees and -20 degrees, respectively. CFD simulations were in agreement with the experimental results and illustrated shifts in local particle deposition associated with changes in insertion angle. Conclusion: The effect of inhaler orientation at the inhaler-mouth interface on MT aerosol deposition appeared to be dependent on velocity, aerosol size, and formulation. These findings not only demonstrate the need for patient education on correct inhaler orientation, but provide important new methods for those designing new inhalers.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据