4.6 Article

Significant Differences in Debris Captured by the Sentinel Dual-Filter Cerebral Embolic Protection During Transcatheter Aortic Valve Replacement Among Different Valve Types

期刊

JACC-CARDIOVASCULAR INTERVENTIONS
卷 11, 期 17, 页码 1683-1693

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcin.2018.06.018

关键词

debris; embolic protection; stroke; TAVR

向作者/读者索取更多资源

OBJECTIVES The aim of this study was to evaluate the debris captured by the Claret Sentinel cerebral embolic dual-filter protection device during transfemoral transcatheter aortic valve replacement (TAVR) with different valve types. BACKGROUND Risk for embolization of debris during TAVR may vary by TAVR device. METHODS The filters of 100 consecutive patients were collected and captured debris was analyzed by histopathology and histomorphometry. Three valve types were implanted: the balloon-expandable Edwards SAPIEN 3 (n = 42), the self-expandable Medtronic Evolut R (n = 35), and the mechanically implantable Boston Scientific Lotus (n = 23). RESULTS Among the 3 groups there was no difference in baseline data, including Society of Thoracic Surgeons score for mortality, calcification, or pre-dilation. The type of captured debris did not differ among the 3 valve types in the proximal or distal filter. With the balloon-expandable valve, there were significantly more patients with large debris measuring >= 1,000 mm. The number of particles in the proximal filter was significantly lower with the Lotus (89.8 +/- 106.3) compared with the Evolut R (187.3 +/- 176.9) and Edwards SAPIEN 3 (172.3 +/- 133.5) valves (p = 0.035). Total tissue area in the proximal filter was significantly smaller for the Lotus compared with the other 2 valve types (7.1 +/- 6.3, 20.1 +/- 19.0, and 21.3 +/- 15.1 mm(2); p = 0.0014). In contrast, for the distal filter, there were no differences with respect to valve type for total tissue area, particle size, and number of particles. CONCLUSIONS A significant difference was observed in the size and number of captured tissue particles with the double-filter embolic protection device among different valve types in patients undergoing TAVR. The largest particles were observed in patients treated with a balloon-expandable valve. (c) 2018 by the American College of Cardiology Foundation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据