4.0 Article Proceedings Paper

Comparison of water isotope-ratio determinations using two cavity ring-down instruments and classical mass spectrometry in continuous ice-core analysis

期刊

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/10256016.2013.781598

关键词

cavity ring-down; hydrogen-2; ice-core; isotope measurements; methods and equipment; oxygen-18; paleoclimatology; water

资金

  1. Office of Polar Programs (OPP)
  2. Directorate For Geosciences [1023672] Funding Source: National Science Foundation
  3. Office of Polar Programs (OPP)
  4. Directorate For Geosciences [1023318] Funding Source: National Science Foundation
  5. Office Of The Director
  6. Office Of Internatl Science &Engineering [0968391] Funding Source: National Science Foundation

向作者/读者索取更多资源

We present a detailed comparison between subsequent versions of commercially available wavelength-scanned cavity ring-down water isotope analysers (L2120-i and L2130-i, Picarro Inc.). The analysers are used in parallel in a continuous mode by adaption of a low-volume flash evaporation module. Application of the analysers to ice-core analysis is assessed by comparison between continuous water isotope measurements of a glacial ice-core from Severnaya Zemlya with discrete isotope-ratio mass spectrometry measurements performed on parallel samples from the same ice-core. The great advances between instrument versions, particularly in the measurement of H-2, allow the continuous technique to achieve the same high level of accuracy and precision obtained using traditional isotope spectrometry techniques in a fraction of the experiment time. However, when applied to continuous ice-core measurements, increased integration times result in a compromise of the achievable depth resolution of the ice-core records.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据