Quantification of Hepatic Steatosis With Dual-Energy Computed Tomography Comparison With Tissue Reference Standards and Quantitative Magnetic Resonance Imaging in the ob/ob Mouse

期刊

INVESTIGATIVE RADIOLOGY
卷 47, 期 10, 页码 603-610

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/RLI.0b013e318261fad0

关键词

dual-energy CT; hepatic steatosis; chemical shift; magnetic resonance imaging; proton density fat-fraction

资金

  1. UW Radiology RD
  2. National Institutes of Health [R01 DK083380, R01 DK088925, RC1 EB010384, T32 CA009206]
  3. Coulter Foundation
  4. GE Healthcare
  5. National Institutes of Health

向作者/读者索取更多资源

Objective: The aim of this study was to compare dual-energy computed tomography (DECT) and magnetic resonance imaging (MRI) for fat quantification using tissue triglyceride concentration and histology as references in an animal model of hepatic steatosis. Materials and Methods: This animal study was approved by our institution's Research Animal Resource Center. After validation of DECT and MRI using a phantom consisting of different triglyceride concentrations, a leptin-deficient obese mouse model (ob/ob) was used for this study. Twenty mice were divided into 3 groups based on expected levels of hepatic steatosis: low (n = 6), medium (n = 7), and high (n = 7) fat. After MRI at 3 T, a DECT scan was immediately performed. The caudate lobe of the liver was harvested and analyzed for triglyceride concentration using a colorimetric assay. The left lateral lobe was also extracted for histology. Magnetic resonance imaging fat-fraction (FF) and DECT measurements (attenuation, fat density, and effective atomic number) were compared with triglycerides and histology. Results: Phantom results demonstrated excellent correlation between triglyceride content and each of the MRI and DECT measurements (r(2) >= 0.96, P <= 0.003). In vivo, however, excellent triglyceride correlation was observed only with attenuation (r(2) = 0.89, P < 0.001) and MRI-FF (r(2) = 0.92, P < 0.001). Strong correlation existed between attenuation and MRI-FF (r(2) = 0.86, P < 0.001). Nonlinear correlation with histology was also excellent for attenuation and MRI-FF. Conclusions: Dual-energy computed tomography (CT) data generated by the current Gemstone Spectral Imaging analysis tool do not improve the accuracy of fat quantification in the liver beyond what CT attenuation can already provide. Furthermore, MRI may provide an excellent reference standard for liver fat quantification when validating new CT or DECT methods in human subjects.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据