4.0 Article

Comparison of 3 Kidney Injury Multiplex Panels in Rats

期刊

INTERNATIONAL JOURNAL OF TOXICOLOGY
卷 31, 期 6, 页码 529-536

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/1091581812463348

关键词

kidney injury biomarkers; multiplex; para-aminophenol; rat nephrotoxicity biomarkers

向作者/读者索取更多资源

Kidney injury biomarkers have been utilized by pharmaceutical companies as a means to assess the potential of candidate drugs to induce nephrotoxicity. Multiple platforms and assay methods exist, but the comparison of these methods has not been described. Millipore's Kidney Toxicity panel, EMD/Novagen's Widescreen Kidney Toxicity panel, and Meso Scales Kidney Injury panel were selected based on published information. Kidney injury molecule 1, cystatin C, clusterin, and osteopontin were the 4 biomarkers common among all kits tested and the focus of this study. Rats were treated with a low and high dose of para-aminophenol, a known nephrotoxicant, and urine samples were collected and analyzed on the Bio-Plex 200 or MSD's Sector Imager 6000, according to manufacturers specifications. Comparatively, of the 3 kits, Millipore was the most consistent in detecting elevations of 3 out of the 4 biomarkers at both dose levels and indicated time points.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据