4.3 Article

Middle temperature scale for infrared radiation thermometer calibrated against multiple fixed points

期刊

INTERNATIONAL JOURNAL OF THERMOPHYSICS
卷 29, 期 3, 页码 1014-1025

出版社

SPRINGER/PLENUM PUBLISHERS
DOI: 10.1007/s10765-008-0403-3

关键词

infrared radiation thermometer; InSb detector; middle temperature; multiple fixed points; temperature scale

向作者/读者索取更多资源

The NMIJ has established a temperature scale for infrared radiation thermometry from 100 to 420 degrees C consisting of three fixed-point blackbodies at the In, Sn, and Zn points and an InSb detector thermometer at a wavelength of 4.7 mu m. The blackbody cavities have large openings of 15 mm diameter. The expanded uncertainties (k = 2) of the In, Sn, and Zn fixed-point radiance temperatures are estimated to be 0.03, 0.03, and 0.05 K, respectively. The expanded uncertainties (k = 2) for the calibration of the infrared radiation thermometer are estimated to be 0.04 K at 157 degrees C (In point), 0.04 K at 232 degrees C (Sn point), and 0.07 K at 420 degrees C (Zn point).