4.6 Article

Mechanics of stretchable electronics on balloon catheter under extreme deformation

期刊

INTERNATIONAL JOURNAL OF SOLIDS AND STRUCTURES
卷 51, 期 7-8, 页码 1555-1561

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.ijsolstr.2014.01.008

关键词

Stretchable electronics; Balloon catheter; Finite deformation; Analytic model

资金

  1. NSF
  2. NSFC
  3. Div Of Civil, Mechanical, & Manufact Inn
  4. Directorate For Engineering [1400169] Funding Source: National Science Foundation

向作者/读者索取更多资源

Stretchable electronics has been applied to balloon catheters for high-efficacy ablation, with tactile sensing integrated on the surface, to establish full and conformal contact with the endocardial surface for elimination of the heart sink caused by blood flow around their surfaces. The balloon of the catheter folds into uniform 'clover' patterns driven by the pressure mismatch inside (similar to vacuum) and outside of the balloon (pressure similar to 1 atm). The balloon catheter, on which microelectrodes and interconnects are printed, undergoes extreme mechanical deformation during its inflation and deflation. An analytic solution is obtained for balloon catheter inflation and deflation, which gives analytically the distribution of curvatures and the maximum strain in the microelectrodes and interconnects. The analytic solution is validated by the finite element analysis. It also accounts for the effect of inflated radius, and is very useful to the optimal design of balloon catheter. Crown Copyright (C) 2014 Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据