Commissioning and quality assurance of RapidArc radiotherapy delivery system

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ijrobp.2008.05.060

关键词

RapidArc; commissioning; QA

向作者/读者索取更多资源

Purpose: The Varian RapidArc is a system for intensity-modulated radiotherapy (IMRT) treatment planning and delivery. RapidArc incorporates capabilities such as variable dose-rate, variable gantry speed, and accurate and fast dynamic multileaf collimators (DMLC), to optimize dose conformality, delivery efficiency, accuracy and reliability. We developed RapidArc system commissioning and quality assurance (QA) procedures. Methods and Materials: Tests have been designed that evaluate RapidArc performance in a stepwise manner. First, the accuracy position during gantry rotation is examined. Second, the ability to vary and control the dose-rate and gantry speed is evaluated. Third, the combined use of variable DMLC speed and dose-rate is studied. Results: Adapting the picket fence test for RapidArc, we compared the patterns obtained with stationary gantry and in RapidArc mode, and showed that the effect of gantry rotation on leaf accuracy was minimal (<= 0.2 mm). We then combine different dose-rates (111-600 MU/min), gantry speeds (5.5-4.3 degrees/s), and gantry range (Delta theta = 90-12.9 degrees) to give the same dose to seven parts of a film. When normalized to a corresponding open field (to account for flatness and asymmetry), the dose of the seven portions show good agreement, with a mean deviation of 0.7%. In assessing DMLC speed (0.46, 0.92, 1.84, and 2.76 cm/s) during RapidArc, the analysis of designed radiation pattern indicates good agreement, with a mean deviation of 0.4%. Conclusions: The results of these tests provide strong evidence that DMLC movement, variable dose-rates and gantry speeds can be precisely controlled during RapidArc. (C) 2008 Elsevier Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据