4.7 Article

Application of process analytical technology in tablet process development using NIR spectroscopy: Blend uniformity, content uniformity and coating thickness measurements

期刊

INTERNATIONAL JOURNAL OF PHARMACEUTICS
卷 357, 期 1-2, 页码 108-118

出版社

ELSEVIER
DOI: 10.1016/j.ijpharm.2008.01.062

关键词

process analytical technology; blend uniformity; NIR; tablet coating; content uniformity

向作者/读者索取更多资源

Near-infrared (NIR)spectroscopy was employed as a process analytical technique in three steps of tabletting process: to monitor the blend homogeneity, evaluate the content uniformity of tablets and determine the tablets coating thickness. A diode-array spectrometer mounted on a lab blender (SP15 NIR lab blender) was used to monitor blend uniformity using a calibration-free model with drug concentration ranging from 2.98 to 9.25% (w/w). The method developed accurately depicted the changes in concentration of the drug during blending and the positive effect of a delumping step in the production process. Blend homogeneity was reached within 2 min of the blending step post-delumping, with relative standard deviation (R.S.D.) values varying from 1.0 to 2.5% depending on the drug concentration of the blend. A Fourier-transform spectrometer (Bruker MPA) was used to analyze content uniformity and coating thickness with calibration based models. Prediction of a validation set with tablets compacted at pressures not present in the calibration set yielded an root mean square error of cross validation (RMSEP) of 1.94%; prediction of tablets compacted at pressures present in the calibration set yielded a RMSEP of 1.48%. Performance of the model was influenced by several physical tablet properties, which could be reduced by spectral pre-processing. A model based on reflectance spectra predicted coating thickness and its variation more accurately than the model based on transmission spectra. Inter-tablet coating variation was predicted with NIR and compared to reference thickness measurements. Both methods gave comparable results. Initial inter-tablet variation of tablets sampled in-process during coating was high, but stabilized after 30 min into the process. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据