4.7 Article

EFFICACY OF A NEW CHARGE-BALANCED BIPHASIC ELECTRICAL STIMULUS IN THE ISOLATED SCIATIC NERVE AND THE HIPPOCAMPAL SLICE

期刊

出版社

WORLD SCIENTIFIC PUBL CO PTE LTD
DOI: 10.1142/S0129065712500311

关键词

Deep brain stimulation; charge-balanced biphasic pulses; peripheral nerve; hippocampal slice

向作者/读者索取更多资源

Most deep brain stimulators apply rectangular monophasic voltage pulses. By modifying the stimulus shape, it is possible to optimize stimulus efficacy and find the best compromise between clinical effect, minimal side effects and power consumption of the stimulus generator. In this study, we compared the efficacy of three types of charge-balanced biphasic pulses (CBBPs, nominal duration 100 mu s) in isolated sciatic nerves and in in vitro hippocampal brain slices of the rat. Using these two models, we tested the efficacy of several stimulus shapes exclusively on axons (in the sciatic nerve) and compared the effect with that of stimuli in the more complex neuronal network of the hippocampal slice by considering the stimulus-response relation. We showed that (i) adding an interphase gap (IPG, range 100-500 mu s) to the CBBP enhances stimulus efficacy in the rat sciatic nerve and (ii) that this type of stimuli (CBBP with IPG) is also more effective in hippocampal slices. This benefit was similar for both models of voltage and current stimulation. In our two models, asymmetric CBBPs were less beneficial. Therefore, CBBPs with IPG appear to be well suited for application to DBS, since they enhance efficacy, extend battery life and potentially reduce harmful side effects.