4.5 Article

Comparison of the biomechanical effects of pertrochanteric fixator and dynamic hip screw on an intertrochanteric femoral fracture using the finite element method

出版社

WILEY
DOI: 10.1002/rcs.1584

关键词

Intertrochanteric fractures; devices; external fixator; dynamic hip screw

类别

向作者/读者索取更多资源

BackgroundThis study was conducted to examine and compare the pertrochanteric fixator (PTF) and dynamic hip screw (DHS). MethodsAnsysWorkbench, a commercial finite element based program, was used to investigate the biomechanical parameters of femoral fractures and implants. The human femoral model was scanned and a femoral intertrochanteric fracture was created. Von Mises stress, von Mises strain, and shear and deformation forces on the upper and lower parts of the fracture line; and the von Mises stress, von Mises strain parameters of the screws and implant bodies during axial loading were calculated and compared. ResultsIt was determined that the magnitude of displacement forces along the fracture line was lower in PTF. Regarding the von Mises stress and von Mises strain values on the implant bodies, the DHS body had higher levels compared with the PTF body. ConclusionThis study demonstrated that PTF is more favorable than DHS for early axial loading and ambulation. Copyright (c) 2014 John Wiley & Sons, Ltd.