Pressurized sample infusion (PSI) continuously delivers solution from a flask through capillary tubing at a flow rate that makes it suitable as a low internal volume pumping method for electrospray ionization mass spectrometry (ESI-MS). The flow rate can be predicted from the applied pressure by using the Hagen-Poiseuille equation, but variability in internal diameter of the PEEK tubing used is such that each individual length should be calibrated if accurate results are sought. Once calibrated, the values hold well for different solvents across a range of pressures. The technique has been exemplified in two reactions: the deprotection of a protected amino acid using trifluoroacetic acid, and in the platinum-catalyzed redistribution of two silanes. In both cases, the abundances of starting material, product(s) and intermediates were tracked in real time as the reaction proceeded. (c) 2012 Elsevier B.V. All rights reserved.