INTERNATIONAL JOURNAL OF MASS SPECTROMETRY
卷 325, 期 -, 页码 67-72出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.ijms.2012.06.007
关键词
FT-ICR; FTMS; Baseline roll; Isotopic distribution; Mass accuracy
资金
- NSF Division of Materials Research [DMR-0654118]
- State of Florida
An absorption-mode Fourier transform ion cyclotron resonance (FT-ICR) mass spectrum exhibits two types of baseline distortion: a slow periodic oscillation even where no signal peaks are present, and additional distortion near signal peaks and proportional to signal magnitude. These distortions interfere with automated peak-picking, unless the baseline systematic variation is much less than baseline rms random noise. We previously showed that the slow oscillation is removed by low-pass filtering. Here, we present a fast, robust, and automated algorithm that flattens the absorption-mode spectral baseline, even in the vicinity of signal peaks. The method begins by defining baseline data values, followed by linear interpolation to generate baseline data values between the defined values, then boxcar smoothing to generate a final baseline spectrum, and final subtraction of that baseline from the original spectrum to yield a baseline-flattened absorption-mode spectrum. We apply the algorithm to a crude oil spectrum (with 8000 peaks) and to a ribonuclease A protein spectrum (with multiply-charged ion isotopic distributions). We identify many more peaks (crude oil) without loss of mass accuracy, and obtain more accurate isotopic distributions (RNase A). (C) 2012 Elsevier B.V. All rights reserved.