4.3 Article

Isotope ratio measurements of pg-size plutonium samples using TIMS in combination with multiple ion counting and filament carburization

期刊

INTERNATIONAL JOURNAL OF MASS SPECTROMETRY
卷 279, 期 2-3, 页码 87-92

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.ijms.2008.10.014

关键词

TIMS; Multiple ion counting; Filament carburization; Multi-dynamic measurement

向作者/读者索取更多资源

A sample preparation procedure for isotopic measurements using thermal ionization mass spectrometry (TIMS) was developed which employs the technique of carburization of rhenium filaments. Carburized filaments were prepared in a special vacuum chamber in which the filaments were exposed to benzene vapour as a carbon supply and carburized electrothermally. To find the optimal conditions for the carburization and isotopic measurements using TIMS, the influence of various parameters such as benzene pressure, carburization current and the exposure time were tested. As a result. carburization of the filaments improved the overall efficiency by one order of magnitude. Additionally, a new multi-dynamic measurement technique was developed for Pu isotope ratio measurements using a multiple ion counting (MIC) system. This technique was combined with filament carburization and applied to the NBL-137 isotopic standard and samples of the NUSIMEP 5 inter-laboratory comparison campaign, which included certified Plutonium materials at the ppt-level. The multi-dynamic measurement technique for Plutonium, in combination with filament carburization, has been shown to significantly improve the precision and accuracy for isotopic analysis of environmental samples with low-levels of plutonium. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据