Localised drug release using MRI-controlled focused ultrasound hyperthermia

期刊

INTERNATIONAL JOURNAL OF HYPERTHERMIA
卷 27, 期 2, 页码 156-171

出版社

TAYLOR & FRANCIS LTD
DOI: 10.3109/02656736.2010.518198

关键词

feedback control; heat targeted drug delivery; high intensity focused ultrasound; MRI thermometry; thermosensitive liposomes

资金

  1. National Cancer Institute of Canada through Terry Fox Foundation
  2. Government of Ontario, Canada

向作者/读者索取更多资源

Materials and methods: Results are reported from ten rabbits, where a FUS beam was scanned in a circular trajectory to heat 10--15 mm diameter regions in normal thigh to 43 degrees A degrees C for 20--30 min. MRI thermometry was used for closed-loop feedback control to achieve temporally and spatially uniform heating. Lyso-thermosensitive liposomal doxorubicin was infused intravenously during hyperthermia. Unabsorbed liposomes were flushed from the vasculature by saline perfusion 2 h later, and tissue samples were harvested from heated and unheated thigh regions. The fluorescence intensity of the homogenised samples was used to calculate the concentration of doxorubicin in tissue. Results: Closed-loop control of FUS heating using MRI thermometry achieved temperature distributions with mean, T90 and T10 of 42.9 degrees A degrees C, 41.0 degrees A degrees C and 44.8 degrees A degrees C, respectively, over a period of 20 min. Doxorubicin concentrations were significantly higher in tissues sampled from heated than unheated regions of normal thigh muscle (8.3 versus 0.5 ng/mg, mean per-animal difference == 7.8 ng/mg, P < 0.05, Wilcoxon matched pairs signed rank test). Conclusions: The results show the potential of MRI-controlled focused ultrasound hyperthermia for enhanced local drug delivery with temperature-sensitive drug carriers.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据