Bone conduction hearing sensitivity in normal-hearing subjects: Transcutaneous stimulation at BAHA vs BCI position

期刊

INTERNATIONAL JOURNAL OF AUDIOLOGY
卷 53, 期 6, 页码 360-369

出版社

TAYLOR & FRANCIS LTD
DOI: 10.3109/14992027.2014.880813

关键词

Bone conduction; hearing sensitivity; transcutaneous stimulation; bone conduction implant; bone-anchored hearing aid; threshold shift; ear-canal sound pressure; objective measure; normal-hearing subjects

资金

  1. Vinnova (Swedish Innovation Agency)

向作者/读者索取更多资源

Objective: Bone conduction (BC) stimulation closer to the cochlea has previously been shown to give higher cochlear promontory acceleration measured by laser Doppler vibrometry (LDV). This study is investigating whether stimulation closer to the cochlea also gives improved hearing sensitivity. Furthermore, the study compares shifts in hearing sensitivity (BC thresholds) and ear-canal sound pressure (ECSP). Design: BC hearing thresholds and ECSP have been measured for stimulation at two positions: the existing bone-anchored hearing aid (BAHA) position, and a new bone conduction implant (BCI) position that is located closer to the cochlea. Study sample: The measurements were made on 20 normal-hearing subjects. Results: Depending on frequency, the ipsilateral hearing threshold was 3-14 dB better, and the ipsilateral ECSP was 2-12 dB higher for the BCI than for the BAHA position, with no significant differences between threshold and ECSP shifts at group level for most frequencies, and individually only for some subjects. Conclusions: It was found that both the objective ECSP and the subjective hearing threshold measurements gave similar improvement as previous LDV measurements for stimulation closer to the cochlea. One exception was that the LDV measurements did not show the improved sensitivity for frequencies below 500 Hz found here.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据