4.5 Article

Few Outflow Problems With a Self-locating Catheter for Peritoneal Dialysis A Randomized Trial

期刊

MEDICINE
卷 94, 期 48, 页码 -

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/MD.0000000000002083

关键词

-

资金

  1. Vasterbotten Kidney Foundation
  2. Vasterbotten County Council, Sweden

向作者/读者索取更多资源

We developed a technique for direct start of peritoneal dialysis. Using a coiled or straight Tenckhoff catheter often results in obstruction of flow. A self-locating Wolfram catheter is on the market. It is not clarified if this results in a benefit.The primary aim of this study was to perform a randomized investigation to clarify if the use of a self-locating peritoneal dialysis (PD) catheter would result in different flow problems than a straight Tenckhoff catheter.A total of 61 insertions were made who were randomized and received either a straight Tenckhoff (n=32) or a self-locating Wolfram catheter (n=29).A previously described operation technique allowed immediate postoperative start of dialysis. Seven straight Tenckhoff catheters had to be changed into self-locating catheters, and none vice versa, due to flow problems (P=0.011). An early leakage resulted in temporarily postponed PD in 4 patients.This study showed that using the present operation technique the self-locating PD-catheter causes fewer obstruction episodes than a straight Tenckhoff catheter. This facilitates immediate postoperative start of PD.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据