4.3 Article

Evaluation of a frequency transposition algorithm using wearable hearing aids

期刊

INTERNATIONAL JOURNAL OF AUDIOLOGY
卷 48, 期 6, 页码 384-393

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/14992020902803138

关键词

Transposition; Hearing loss; Dead region; Hearing aid; Consonant recognition

资金

  1. RNID
  2. MRC
  3. Medical Research Council [G0701870] Funding Source: researchfish
  4. MRC [G0701870] Funding Source: UKRI

向作者/读者索取更多资源

Transposition of acoustic information from higher to lower frequencies may help people with severe or profound high-frequency hearing loss, especially when a 'dead region' is present. Previously, we (Robinson et al. 2007) evaluated the benefit of an FFT-based transposition algorithm in a laboratory study. Although results were promising, we hypothesized that further training and exposure would be needed to gain the full benefit. This was tested here by implementing the algorithm in wearable digital hearing aids. Five subjects with high-frequency dead regions used the aids for five weeks. Performance on the transposing and control conditions was compared objectively using speech tests (vowel-consonant-vowel, 's' detection, and speech in noise) and subjectively using questionnaires. Overall, the results showed no benefit with the transposition even after experience. Subjective preference was generally for the control condition.