4.6 Article

A 32 x 32 Element Row-Column Addressed Capacitive Micromachined Ultrasonic Transducer

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TUFFC.2011.1937

关键词

-

资金

  1. Natural Sciences and Engineering Research Council of Canada
  2. Canadian Institutes of Health Research
  3. Waterloo Institute of Nanotechnology
  4. University of Waterloo
  5. National Science Foundation [ECS-0335765]

向作者/读者索取更多资源

This paper presents characterization and initial imaging results of a 32 x 32 element two-dimensional capacitive micromachined ultrasonic transducer array. The devices are fabricated using a wafer bonding process in which both the insulation layer and the membrane are user-deposited silicon nitride. The transducers use a row-column addressing scheme to simplify the fabrication process and beamformer. By adjusting the number of rows and columns that are biased, the effective aperture of the transducer can be adjusted. This is significant because it permits imaging in the near-field of the transducer without the use of a lens. The effect on the transmit beam profile is demonstrated. The transducer has a center frequency of 5.9 MHz and a relative bandwidth of 110%. Images of horizontal and vertical wires are taken to demonstrate image resolution. A three-dimensional image of four pin heads is also demonstrated.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据