期刊
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL
卷 55, 期 6, 页码 1394-1399出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TUFFC.2008.804
关键词
-
资金
- NIBIB NIH HHS [P41 EB002182, P41-EB2182, P41 EB002182-12] Funding Source: Medline
- NATIONAL INSTITUTE OF BIOMEDICAL IMAGING AND BIOENGINEERING [P41EB002182] Funding Source: NIH RePORTER
A high-frequency angled needle ultrasound transducer with an aperture size of 0.4 x 0.56 mm(2) was fabricated using a lead zinc niobate-lead titanate (PZN-7%PT) single crystal as the active piezoelectric material. The single crystal was bonded to a conductive silver particle matching layer and a conductive epoxy backing material through direct contact curing. A parylene outer matching layer was formed by vapor deposition. Angled needle probe configuration was achieved by dicing at 451 to the single crystal poling direction to satisfy a clinical request for blood flow measurement in the posterior portion of the eye. The electrical impedance magnitude and phase of the transducer were 42 Omega and -63 degrees, respectively. The measured center frequency and the fractional bandwidth at -6 dB were 43 MHz and 45%, respectively. The two-way insertion loss was approximately 17 dB. Wire phantom imaging using fabricated PZN-7%PT single crystal transducers was obtained and spatial resolutions were assessed.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐