4.5 Article Proceedings Paper

Fabrication Methodology of Enhanced Stability Room Temperature TlBr Gamma Detectors

期刊

IEEE TRANSACTIONS ON NUCLEAR SCIENCE
卷 60, 期 2, 页码 1231-1236

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2013.2252363

关键词

Room temperature gamma detector; semiconductor radiation detector; thallium bromide (TlBr)

向作者/读者索取更多资源

Thallium bromide (TlBr) is a material of interest for use in room temperature gamma ray detector applications due to is wide bandgap 2.7 eV and high average atomic number (Tl81, Br 35). Researchers have achieved energy resolutions of 1.3% at 662 keV, demonstrating the potential of this material system. However, these detectors are known to polarize using conventional configurations, limiting their use. While high quality material is a critical starting point for excellent detector performance, we show that the room temperature stability of planar TlBr gamma spectrometers can be significantly enhanced by treatment with both hydrofluoric and hydrochloric acid. By incorporating F or Cl into the surface of TlBr, current instabilities are eliminated and the longer term current of the detectors remains unchanged. In addition the choice of electrode metal is shown to have a dramatic effect on the long term stability of TlBr detector performance Am-241 spectra are also shown to be more stable for extended periods; detectors have been held at 4000 V/cm for 50 days with less than 10% degradation in peak centroid position.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据