IEEE TRANSACTIONS ON NUCLEAR SCIENCE
卷 60, 期 5, 页码 3172-3180出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2013.2280631
关键词
Detector instrumentation; PET
资金
- Japan Science and Technology Agency (JST), Development of Systems and Technology for Advanced Measurement and Analysis
The X'tal cube is our original PET detector, which is being developed to achieve isotropic 3D positioning detectability. The X'tal cube is based on a 3D segmented crystal block for which all surfaces are covered with photo-detectors. Previously, instead of our initial approach of gluing segmented pieces of crystals, we successfully constructed a crystal block segmented by laser processing, and we developed the X'tal cube with the laser-processed 3D square grids of 2 mm length. In this paper, we extend the laser processing to 3D square grids of 1 mm length. The volume of a 1-mm crystal segment is 1/8 of that of a 2-mm crystal segment. We also evaluate imaging resolution performance with a newly developed one-pair prototype system to simulate a ring-type scanner, while our previous reports were limited to development and evaluation of a single detector. In particular, we compare 1-mm X'tal cube detectors to 2-mm X'tal cube detectors. The one-pair prototype system consisted of two X'tal cubes, two rotating stages, and a 192-channel data acquisition system. Each X'tal cube consisted of the LYSO cubic crystal block of 18 x 18 x 18 mm(3) in which the 3D grids of 1 mm pitch were fabricated by internal laser processing. The 4 x 4 arrays of multi pixel photon counters were optically coupled to each surface of the crystal block. The detector positions were automatically controlled to simulate a ring-type PET with a 14.6 cm diameter. Data were collected for all assumed detector positions and then a sinogram was obtained. The data were reconstructed using filtered backprojection. The average spatial resolution of the 2-mm X'tal cube was 1.9 mm full width at half maximum (FWHM) over the field-of-view (FOV). On the other hand, the average spatial resolution of the 1-mm X'tal cube was 1.3 mm FWHM over the FOV. By applying deconvolution with the assumption that the point source was a Gaussian function of 1.0 mm FWHM, we estimated the average spatial resolution of the 2-mm X'tal cube and the 1-mm X'tal cube as 1.6 mm FWHM and 0.83 mm FWHM, respectively. The average spatial resolution of the 1-mm X'tal cube was improved approximately 48% compared to the 2-mm X'tal cube. In conclusion, we confirmed the potential of the X'tal cube for uniform and high resolution imaging.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据