4.5 Article

Positron Emission Tomography-Guided Biopsy With a Dedicated Breast Scanner: Initial Evaluation

期刊

IEEE TRANSACTIONS ON NUCLEAR SCIENCE
卷 56, 期 3, 页码 620-624

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2009.2015449

关键词

Breast imaging; nuclear medicine; instrumentation; biopsy

资金

  1. National Cancer Institute [R01CA094196]

向作者/读者索取更多资源

Efforts to create dedicated nuclear medicine breast imaging scanners are maturing as they approach more routine clinical use. These systems promise to improve the detection of small lesions, especially those in breasts that are not definitively imaged with standard methods. Since these techniques lack sufficient specificity for accurate diagnoses of cancer, a method for guiding biopsies with these new devices must be developed. In this investigation, a technique that capitalizes upon the three-dimensional, multi-angle positron emission tomography (PET) breast images created by our new positron emission mammography/tomography imaging and biopsy device (PEM-PET) to guide breast biopsies was tested. Specifically, the accuracy and precision of the needle positioning system was measured for each of the three axes of motion. In addition, a biopsy of a simulated breast lesion was performed under realistic imaging conditions. The results from the positioning tests revealed that the system had very good accuracy and precision in all three axes. The maximum positioning error was +/- 0.27 mm, sufficient to obtain samples from most lesions detected with PEM-PET. In addition, the simulated biopsy was performed successfully on a single attempt in obtaining a sample from a 5-mm-diameter target sphere. This successful test moves the project closer to the next phase of development, initiation of limited human testing.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据