Performance Characteristics of a New LSO PET/CT Scanner With Extended Axial Field-of-View and PSF Reconstruction

期刊

IEEE TRANSACTIONS ON NUCLEAR SCIENCE
卷 56, 期 3, 页码 633-639

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2009.2015764

关键词

Extended FOV; NEMA measurements; PET/CT; PSF reconstruction

资金

  1. National Institute of Health (NIH) [CA 94317]

向作者/读者索取更多资源

A new combined lutetium oxyorthosilicate (LSO) PET/CT scanner with an extended axial field-of-view (FOV) of 21.8 cm has been developed (Biograph TruePoint PET/CT with TrueV; Siemens Molecular Imaging) and introduced into clinical practice. The scanner includes the recently announced point spread function (PSF) reconstruction algorithm. The PET components incorporate four rings of 48 detector blocks, 5.4 cm x 5.4 cm in cross-section. Each block comprises a 13 x 13 matrix of 4 x 4 x 20 mm(3) elements. Data are acquired with a 4.5 ns coincidence time window and an energy window of 425-650 keV. The physical performance of the new scanner has been evaluated according to the recently revised National Electrical Manufacturers Association (NEMA) NU 2-2007 standard and the results have been compared with a previous PET/CT design that incorporates three rings of block detectors with an axial coverage of 16.2 cm (Biograph TruePoint PET/CT; Siemens Molecular Imaging). In addition to the phantom measurements, patient Noise Equivalent Count Rates (NECRs) have been estimated for a range of patients with different body weights (42-154 kg). The average spatial resolution is the same for both scanners: 4.4 mm (FWHM) and 5.0 mm (FWHM) at 1 cm and 10 cm respectively from the center of the transverse FOV. The scatter fractions of the Biograph TruePoint and Biograph TruePoint TrueV are comparable at 32%. Compared to the three ring design, the system sensitivity and peak NECR with smoothed randoms correction (1R) increase by 82% and 73%, respectively. The increase in sensitivity from the extended axial coverage of the Biograph TruePoint PET/CT with TrueV should allow a decrease in either scan time or injected dose without compromising diagnostic image quality. The contrast improvement with the PSF reconstruction potentially offers enhanced detectability for small lesions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据