4.7 Article

Unmixing Molecular Agents From Absorbing Tissue in Multispectral Optoacoustic Tomography

期刊

IEEE TRANSACTIONS ON MEDICAL IMAGING
卷 33, 期 1, 页码 48-60

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TMI.2013.2279994

关键词

Molecular imaging; multispectral imaging; optoacoustic/photoacoustic tomography; spectral unmixing; sub-pixel target detection

资金

  1. DFG [GRK 1371]
  2. European Research Council through the Advanced Investigator Award
  3. European Union project FAMOS (FP7 ICT) [317744]

向作者/读者索取更多资源

Detection of intrinsic or extrinsically administered chromophores and photo-absorbing nanoparticles has been achieved by multi-spectral optoacoustic tomography (MSOT). The detection sensitivity of MSOT depends not only on the signal to noise ratio considerations, as in conventional optoacoustic (photoacoustic) tomography implementations, but also on the ability to resolve the molecular targets of interest from the absorbing tissue background by means of spectral unmixing or sub-pixel detection methods. However, it is not known which unmixing methods are optimally suited for the characteristics of multispectral optoacoustic images. In this work we investigated the performance of different sub-pixel detection methods, typically used in remote sensing hyperspectral imaging, within the context of MSOT. A quantitative comparison of the different algorithmic approaches was carried out in an effort to identify methods that operate optimally under the particulars of molecular imaging applications. We find that statistical sub-pixel detection methods can demonstrate a unique detection performance with up to five times enhanced sensitivity as compared to linear unmixing approximations, under the condition that the optical agent of interest is sparsely present within the tissue volume, as common when using targeted agents and reporter genes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据