4.7 Article

A New Impedance Measurement System for PZT-Based Structural Health Monitoring

期刊

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TIM.2009.2018693

关键词

Impedance analyzer; smart structures; structural health monitoring (SHM); virtual instrumentation

资金

  1. Brazil's Agency for the Financing Studies and Projects (FINEP)
  2. Brazil's National Council for Scientific and Technological Development (CNPq)

向作者/读者索取更多资源

Recent research has shown that electromechanical (E/M) impedance-based techniques are an efficient strategy for damage sensing in the context of structural health monitoring (SHM). The basic principle is to excite the lead zirconate titanate (PZT) sensors attached to the structure and to measure the corresponding electrical impedance. A variation in the electrical impedance indicates a variation in the mechanical impedance and, consequently, structural damage. To measure the electrical impedance, most universities and research centers use expensive, bulky, heavy, and high-powered impedance analyzers, which limit new developments and practical applications. To overcome these issues, this paper presents an efficient and inexpensive methodology for electrical impedance measurement. To verify the efficiency and accuracy of the new methodology, practical tests were carried out, and the results were compared with those obtained using a conventional impedance analyzer.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据