期刊
IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE
卷 13, 期 1, 页码 111-120出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TITB.2008.2007194
关键词
Ballistocardiography (BCG); home health care; long-term monitoring; sleep
类别
资金
- University of Virginia's General Clinical Research Center [5 M01 RR00847]
A Techniques such as ballistocardiography (BCG) that can provide noninvasive long-term physiological monitoring have gained interest due to a growing recognition of adverse effects from poor sleep and sleep disorders. The noninvasive analysis of physiological signals (NAPS) system is a BCG-based monitoring system developed to measure heart rate, breathing rate, and musculoskeletal movement that shows promise as a general sleep analysis tool. Overnight sleep studies were conducted on 40 healthy subjects during a clinical trial at the University of Virginia. The NAPS system's measures of heart rate and breathing rate were compared to ECG, pulse oximetry, and respiratory inductance plethysmography (RIP). The subjects were split into a training dataset and a validation dataset, maintaining similar demographics in each set. The NAPS system accurately detected heart rate, averaged over the prescribed 30-s epochs, to within less than 2.72 beats per minute of ECG, and accurately detected breathing rate, averaged over the same epochs, to within 2.10 breaths per minute of RIP hands used in polysomnography.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据