4.6 Article

Multisensor Data Fusion for Physical Activity Assessment

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 59, 期 3, 页码 687-696

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2011.2178070

关键词

Physical activity assessment; sensor fusion; support vector machines; wearable sensing system

资金

  1. NIH [UO1 CA130783]

向作者/读者索取更多资源

This paper presents a sensor fusion method for assessing physical activity (PA) of human subjects, based on support vector machines (SVMs). Specifically, acceleration and ventilation measured by a wearable multisensor device on 50 test subjects performing 13 types of activities of varying intensities are analyzed, from which activity type and energy expenditure are derived. The results show that the method correctly recognized the 13 activity types 88.1% of the time, which is 12.3% higher than using a hip accelerometer alone. Also, the method predicted energy expenditure with a root mean square error of 0.42 METs, 22.2% lower than using a hip accelerometer alone. Furthermore, the fusion method was effective in reducing the subject-to-subject variability (standard deviation of recognition accuracies across subjects) in activity recognition, especially when data from the ventilation sensor were added to the fusion model. These results demonstrate that the multisensor fusion technique presented is more effective in identifying activity type and energy expenditure than the traditional accelerometer-alone-based methods.