4.6 Article

In Vivo Validation of a Hybrid Tracking System for Navigation of an Ultrathin Bronchoscope Within Peripheral Airways

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 57, 期 3, 页码 736-745

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2009.2034733

关键词

Bronchoscopy; image-guided surgery; respiratory motion; tracking

资金

  1. National Institutes of Health/National Cancer Institute [R21/R33, CA094303]
  2. PENTAX (HOYA Corporation, Tokyo, Japan)

向作者/读者索取更多资源

Transbronchial biopsy of peripheral lung nodules is hindered by the inability to access lesions endoluminally due to the large diameter of conventional bronchoscopes. An ultrathin scanning fiber bronchoscope has recently been developed to advance image-guided biopsy several branching generations deeper into the peripheral airways. However, navigating a potentially complex 3-D path to the region of interest presents a challenge to the bronchoscopist. An accompanying guidance system has also been developed to track the bronchoscope through the airways, and display its position and intended path on a virtual display. Intraoperative localization of the bronchoscope was achieved by combining electromagnetic tracking (EMT) and image-based tracking (IBT). An error-state Kalman filter was used to model the disagreement between the two tracking sources. The positional tracking error was reduced from 14.22 and 14.92 mm by independent EMT and IBT, respectively, to 6.74 mm using the hybrid approach. Hybrid tracking of the scope orientation and respiratory motion compensation further improved tracking accuracy and stability, resulting in an average tracking error of 3.33 mm and 10.01 degrees.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据