PurposeTo develop a method for spoiling transverse magnetizations without additional gradients to minimize repetition times for radial fast low angle shot (FLASH) MRI. MethodsResidual steady state transverse magnetizations and corresponding image artifacts were analyzed for radial gradient echo sequences with constant and randomized RF phases in comparison with a sequence with refocused frequency-encoding gradients, constant spoiler gradient, and conventional RF spoiling (gold standard). The spoiling performance was assessed for different radial trajectories using numerical simulations, phantom experiments, and in vivo MRI studies of the human brain. ResultsSimulations as well as phantom and in vivo measurements reveal a highly efficient spoiling capacity for randomized RF phases and radial FLASH sequences without the need for gradient rewinding and spoiler gradients. The data also demonstrate a strong dependence of the spoiling performance on the chosen radial trajectory (ie, the azimuthal angular increment between successive projections) with excellent results for an interleaved multiturn scheme. ConclusionEffective spoiling of transverse magnetizations in radial FLASH MRI may be achieved by randomized RF phases without additional spoiler gradients. The technique allows for short repetition times as required for high-speed real-time MRI. Magn Reson Med 75:2094-2099, 2016. (c) 2015 Wiley Periodicals, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据