Optimal acquisition and modeling parameters for accurate assessment of low Ktrans blood-brain barrier permeability using dynamic contrast-enhanced MRI

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 75, 期 5, 页码 1967-1977

出版社

WILEY
DOI: 10.1002/mrm.25793

关键词

K-trans estimation; DCE-MRI; blood-brain barrier; parameter optimization; permeability

资金

  1. NIBIB [EB000993]
  2. Beckman Institute
  3. NINDS [R37NS34467]
  4. NIA [R37AG23084, R01AG039452]

向作者/读者索取更多资源

PurposeTo determine optimal parameters for acquisition and processing of dynamic contrast-enhanced MRI (DCE-MRI) to detect small changes in near normal low blood-brain barrier (BBB) permeability. MethodsUsing a contrast-to-noise ratio metric (K-CNR) for K-trans precision and accuracy, the effects of kinetic model selection, scan duration, temporal resolution, signal drift, and length of baseline on the estimation of low permeability values was evaluated with simulations. ResultsThe Patlak model was shown to give the highest K-CNR at low K-trans. The K-trans transition point, above which other models yielded superior results, was highly dependent on scan duration and tissue extravascular extracellular volume fraction (v(e)). The highest K-CNR for low K-trans was obtained when Patlak model analysis was combined with long scan times (10-30 min), modest temporal resolution (<60 s/image), and long baseline scans (1-4 min). Signal drift as low as 3% was shown to affect the accuracy of K-trans estimation with Patlak analysis. ConclusionDCE acquisition and modeling parameters are interdependent and should be optimized together for the tissue being imaged. Appropriately optimized protocols can detect even the subtlest changes in BBB integrity and may be used to probe the earliest changes in neurodegenerative diseases such as Alzheimer's disease and multiple sclerosis. Magn Reson Med 75:1967-1977, 2016. (c) 2015 Wiley Periodicals, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据