4.7 Article

Respiratory Monitoring System on the Basis of Capacitive Textile Force Sensors

期刊

IEEE SENSORS JOURNAL
卷 11, 期 5, 页码 1112-1119

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JSEN.2010.2082524

关键词

Capacitive force sensors; capacitive measurement; respiration monitoring; textile force sensors

向作者/读者索取更多资源

A novel approach is presented for non-constrictive long-term-monitoring of respiration which could particularly become suitable for home care applications. The system is based on textile integrated force sensors which detect expansion of the thorax during respiration and allow wireless data transmission for maximum mobility. Possible applications include long-term monitoring of patients with chronic pulmonary diseases, early recognition of diseases and the performance measurement of athletes during exercise. Results from performance tests under various conditions are presented. The tests show a high correlation with measurements made with a flow meter as gold standard, allowing conclusions about the respiratory volume. Different respiration patterns are reliably detected and can easily be subdivided even during intensive motion of the person. The sensors' textile buildup may allow a future integration into clothing and thus a minimal constriction for the patient. Nevertheless, the respiratory volume estimations are still not accurate enough mainly because of the use of a simplified volume calculation model and the disregard of the abdominal volume changes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据