Needle enzyme electrode for lactate measurement in vivo

期刊

IEEE SENSORS JOURNAL
卷 8, 期 1-2, 页码 113-120

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JSEN.2007.912919

关键词

animal; biosensors; blood; computer-aided analysis; lactate; measurement; subcutaneous

向作者/读者索取更多资源

Electrochemical lactate needle enzyme electrodes were fabricated based on lactate oxidase with a conventional hydrogen peroxide detection regimen with a linear range up to 7 mM, response time similar to 3 min, and sensitivity similar to 1 nA/mM. A negatively charged inner (sulphonated polyether ether sulphone-polyether sulphone) membrane was applied for ensuring selectivity by limiting oxidazible anion diffusion to the Pt working electrode; polyurethane outer membrane layers were dip coated over the enzyme layer to limit substrate diffusion to the enzyme layer to achieve: 1) stir independence and 2) a low oxygen requirement. Lactate was monitored subcutaneously in rats during controlled haemorrhage and hypovolaemic shock. Tissue lactate showed agreement with blood lactate before haemorrhage and for limited haemorrhage (up to 2 ml blood withdrawal from 16 ml total blood volume) but with blood loss above 3 ml the tissue lactate rise was less pronounced than in blood. Loss of intercompartmental equilibrium due to diffusion limitation is suggested as a factor in causing this difference. An experimental in vitro model was developed which employed the needle lactate electrode within a cylindrical collagen gel to monitor inward diffusion of lactate as a basis for determining lactate diffusion coefficient. The high precision measurement gave a diffusion coefficient consistent with report values 3.54 x 10(-6) cm(2)/s. The simplified experimental approach could allow lactate transport studies across tissue analogues.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据