4.0 Article

Initial clinical experience with a new heparin-coated chronic hemodialysis catheter

期刊

HEMODIALYSIS INTERNATIONAL
卷 13, 期 3, 页码 329-334

出版社

WILEY
DOI: 10.1111/j.1542-4758.2009.00339.x

关键词

Hemodialysis; hemodialysis catheter; catheter infection

向作者/读者索取更多资源

In this paper we wish to report our clinical experience with a new heparin-coated dialysis catheter with a symmetric tip. Over a 16-month period, 60 heparin-coated Tal Palindrome((TM)) catheters were placed in 57 patients. Catheter patency, catheter-related complications, and reasons for catheter removal were recorded. The patient's initial cause of end-stage renal disease, underlying diseases, and site of access were recorded as well. Patients were specifically followed for development of heparin-induced thrombocytopenia. Patient ages were 34-91 (average 66). Fifty-four percent of patients had a history of diabetes. Sixty catheters were placed for a total of 5353 catheter-days. The average catheter indwell time was 107 days (range of 2-381 days). Catheter-related infection occurred in 6 patients over the study period, with a rate of 1.12/1000 catheter-days. Bacteremia occurred in 3 patients with a rate of 0.56/1000 catheter-days. Six catheters were removed or exchanged due to malfunction. There was no incidence of heparin-induced thrombocytopenia. Initial clinical experience with the heparin-coated Tal Palindrome((TM)) hemodialysis catheter demonstrated safe, reliable use, and low infection rates.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据