A novel radiofrequency ablation catheter using contact force sensing: Toccata study

期刊

HEART RHYTHM
卷 9, 期 1, 页码 18-23

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.hrthm.2011.08.021

关键词

Ablation; Atrial fibrillation; Catheter ablation; Supraventricular tachycardia; Contact force

向作者/读者索取更多资源

OBJECTIVES The aim of this multicenter study was to evaluate the device-and procedure-related safety of a novel force-sensing radiofrequency (RF) ablation catheter capable of measuring the real-time contact force (CF) and to present CF data and its possible implications on patient safety. BACKGROUND The clinical outcome of RF ablation for the treatment of cardiac arrhythmias may be affected by the CF between the catheter tip and the tissue. Insufficient CF may result in an ineffective lesion, whereas excessive CF may result in complications. METHODS Seventy-seven patients (43 with right-sided supraventricular tachycardia [SVT] and 34 with atrial fibrillation [AF]) received percutaneous ablation with the novel studied catheter. The CF applied and safety events related to the procedure were reported. RESULTS CF values at mapping ranged from 8 +/- 8 to 60 +/- 35 g and from 12 +/- 10 to 39 +/- 29 g in the SVT group and the LA group, respectively, showing a significant interinvestigator variability (P < .0001). High transient CFs (>100 g) were noted in 27 patients (79%) of the LA group. One device-related complication (tamponade, 3%) occurred in the AF group. CONCLUSIONS Catheter ablation using real-time CF technology is safe for the treatment of SVT and AF. High CFs may occur during catheter manipulation and not just during ablation, suggesting that measuring CF may provide additional useful information to the operator for safe catheter manipulation. In the future, CF-sensing catheters may also increase the effectiveness of RF ablations by allowing better control of the RF lesion size.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据