4.5 Article

New device for single-stage in-office secondary tracheoesophageal puncture: Animal studies

出版社

WILEY-BLACKWELL
DOI: 10.1002/hed.23392

关键词

tracheoesophageal puncture; transnasal esophagoscopy; voice prosthesis; office-based surgery; medical device; laryngectomy

资金

  1. National Medical Research Council of Singapore

向作者/读者索取更多资源

BackgroundA new device is described for secondary tracheoesophageal puncture (TEP). We hypothesize that the device can be used to create the puncture, size, and insert the voice prosthesis in antegrade fashion in a single procedure suitable for office-based use. MethodsA measurement and insertion device (MAID) was designed with the hypothesis in mind and tested on 3 pigs while they were under general anesthesia. ResultsThe procedure was easy and successful in all 3 animals. The MAID allowed measurement of tracheoesophageal wall thickness and antegrade insertion of voice prosthesis with a single device. Transnasal esophagoscopy (TNE) allowed visualization of markings on the MAID within the esophagus to facilitate sizing. Intraesophageal visualization using TNE also enabled accurate positioning, reduced risk of posterior esophageal wall injury, and practically eliminated false passage formation. ConclusionWe believe the MAID can facilitate single-stage office-based secondary TEP in postlaryngectomy patients. (c) 2013 Wiley Periodicals, Inc. Head Neck 36: 882-886, 2014

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据