4.3 Article

Evaluation of the activated partial thromboplastin time assay for clinical monitoring of PEGylated recombinant factor VIII (BAY 94-9027) for haemophilia A

期刊

HAEMOPHILIA
卷 20, 期 4, 页码 593-600

出版社

WILEY
DOI: 10.1111/hae.12374

关键词

aPTT assays; clotting time; coagulation tests; factor VIII; haemophilia; PEGylation

资金

  1. Bayer HealthCare

向作者/读者索取更多资源

Patients with haemophilia (PWH) are usually monitored by the one-stage activated partial thromboplastin time (aPTT) factor VIII (FVIII) assay. Different aPTT activators may affect clotting time (CT) and FVIII:C levels in patients treated with PEGylated FVIII. To evaluate the characteristics of PEGylated FVIII (BAY 94-9027) in various aPTT clotting assays, and to identify suitable aPTT reagents for monitoring BAY 94-9027 during the treatment of PWH, BAY 94-9027 and World Health Organization (WHO) 8th FVIII standards (WHO-8) were spiked into pooled and individual severe haemophilia A plasma at 1.0, 0.25 and 0.05IUmL-1. Five commercial aPTT reagents widely used in clinical laboratories were compared and evaluated for BAY 94-9027 activity in plasma from PWH. BAY 94-9027 and WHO-8 bestowed similar CT and excellent precision when ellagic acid (SynthAFax, Dade Actin, and Cephascreen) aPTT reagents were used. In contrast, BAY 94-9027 showed significantly prolonged CT and poor precision compared with WHO-8 using silica aPTT reagents (APTT-SP and STA PTT 5). Furthermore, free 60-kDa polyethylene glycol (PEG), used for the conjugation of FVIII, showed a dose-dependent prolongation of CT in the APTT-SP assay. There was no effect on the SynthAFax-APTT, prothrombin time, or FXIa-initiated thrombin generation assay, demonstrating that the PEG moiety on FVIII has no general effect on the coagulation cascade. In summary, ellagic aPTT reagents (SynthAFax, Dade Actin, and Cephascreen) are most suitable for evaluating potency of BAY 94-9027 and should be the preferred aPTT reagents used in clinical laboratories for monitoring FVIII activity after infusion of BAY 94-9027 to PWH.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据