4.7 Article

The use of inductively coupled plasma mass spectrometry (ICP-MS) for the determination of toxic and essential elements in different types of food samples

Journal

FOOD CHEMISTRY
Volume 112, Issue 3, Pages 727-732

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.foodchem.2008.06.010

Keywords

ICP-MS; Food samples; Trace elements; Rapid determination

Ask authors/readers for more resources

This paper describes a simple method for the determination of sixteen elements in food samples by using inductively coupled plasma spectrometry (ICP-MS). Prior to analysis, 100-250 mg of powdered food samples were accurately weighed into a Teflon digestion vessel. Then, 4 ml of 20% v/v concentrated nitric acid and 2 ml of hydrogen peroxide were added. Decomposition of samples was carried out in a microwave digestion system. In order to verify the accuracy and precision of the proposed method, five Standard Reference Materials from the National Institute of Standards and Technology (NIST) (Whole Egg Powder RM 8415, Rice Flour SRM 1568a, Typical Diet SRM 1548a, Wheat Flour SRM 1567a and Bovine Muscle Powder RM 8414) were analyzed. Additional validation data are provided based on the analysis of 18 different types of food samples by the proposed method and using comparative methods with AAS as the detector. (c) 2008 Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available